CALIBRATION THERMOCOUPLE K-TYPE ON THE RANGE TEMPERATURE -10°C TO 30°C USING ASTM THERMOMETER AS REFERENCE

Authors

  • Ichsan Purnama Gunadarma University
  • R.R Sri Poernomo Sari Gunadarma University

DOI:

https://doi.org/10.63891/j-mart.v1i1.6

Keywords:

Calibration, Thermocouple, LabView

Abstract

This research focuses on the calibration of four type K thermocouples in the temperature range -10°C to 30°C, using chamber water bath with high precision. The calibration process is carried out by comparing thermocouple readings with readings from ASTM standard thermometers, with data acquisition managed using National Instruments hardware and LabVIEW software. The results of research on the accuracy of type K thermocouples show a shift in linearity and uncertainty values/ uncertainty are within acceptable engineering tolerances. A linear correction equation for each thermocouple is derived to minimize measurement error, ensuring that the thermocouple provides reliable temperature readings for applications within the specified range. This research shows the importance of using high-precision calibration systems, such as ASTM thermometers, to improve thermocouple measurement accuracy, providing a comprehensive solution for precision temperature measurements.

References

[1] Holman, J. P. (2012). Experimental Methods for Engineers (8th ed.). McGraw-Hill.

[2] ASTM E220-19. (2019). Standard Practice for Calibration of Termokopels by Comparison Techniques. ASTM International.

[3] Bentley, R. E. (1998). "Termokopel Temperature Measurement." Measurement Science and Technology, 9(7), 1067–1077.

[4] Chisholm, M. F., & Thompson, A. (2005). "Termokopel Calibration: Precision and Uncertainty Analysis." Journal of Measurement Techniques, 12(4), 245–253.

[5] Smith, J. R., et al. (2021). Stability of Water Bath Chamber for Thermocouple Calibration. Journal of Temperature Measurement, 35(4), 412-420.

[6] Li, Z., & Wang, M. (2020). The Importance of Standard Thermometers in Thermocouple Calibration. International Journal of Metrology, 18(6), 125–132.

[7] Ahmad, S., et al. (2020). Nonlinearity and Material Instability of K-type Termokopels at Low and Medium Temperatures. Journal of Industrial Temperature Control, 45(2), 223–231.

[8] Zhao, L., et al. (2018). Temperature Distribution and Stability in Water Bath Chambers for Thermocouple Calibration. Journal of Experimental Physics, 22(3), 197–206.

[9] Nelson, P., et al. (2022). Accuracy of ASTM Thermometers in Calibration Procedures for High-Precision Temperature Sensors. Temperature Measurement and Control, 59(7), 401–408.

[10] Kumar, R., et al. (2019). Enhancement of Calibration Efficiency Using National Instruments Data Acquisition Systems. Measurement Science & Technology, 30(5), 800-810.

[11] Smith, R. & Brown, H. (2021). Automated Data Acquisition and Control in Temperature Calibration Using LabVIEW. Journal of Applied Instrumentation, 33(8), 1032–1040.

Downloads

Published